Health product recall

B. Braun IV Administration Sets & Infusomat® Space® Administration Sets

Last updated

Summary

Product
B. Braun IV Administration Sets & Infusomat® Space® Administration Sets
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information.

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

Basic Administration Sets With Two Injection Sites

More than 10 numbers, contact manufacturer.

354202

Basic Administration Sets With Two Injection Sites

More than 10 numbers, contact manufacturer.

352604

Basic Administration Sets With Two Injection Sites

More than 10 numbers, contact manufacturer.

354204

Basic Administration Sets With Two Injection Sites

More than 10 numbers, contact manufacturer.

354203

Basic Administration Sets With Two Injection Sites

More than 10 numbers, contact manufacturer.

352049-C

Infusomat® Space® Pump IV Set

More than 10 numbers, contact manufacturer.

490105

Infusomat® Space® Pump IV Set

More than 10 numbers, contact manufacturer.

490101

Infusomat® Space® Pump IV Set

More than 10 numbers, contact manufacturer.

490102

Infusomat® Space® Pump IV Set

More than 10 numbers, contact manufacturer.

490103

Infusomat® Space® Pump IV Set

More than 10 numbers, contact manufacturer.

362033

Infusomat® Space® Pump IV Set

More than 10 numbers, contact manufacturer.

362432

Infusomat® Space® Pump IV Set

More than 10 numbers, contact manufacturer.

490104

Infusomat® Space® Pump IV Set

More than 10 numbers, contact manufacturer.

362032

Infusomat® Space® Pump IV Set

More than 10 numbers, contact manufacturer.

362050

Infusomat® Space® Pump IV Set

More than 10 numbers, contact manufacturer.

490100

IV Administration Set

More than 10 numbers, contact manufacturer.

375100-C

Filtered Administration Sets

More than 10 numbers, contact manufacturer.

354207

Primary ADDitIV® Administration Sets With Check Valve

More than 10 numbers, contact manufacturer.

NF1251

Primary ADDitIV® Administration Sets With Check Valve

More than 10 numbers, contact manufacturer.

V1443

Primay ADDitIV® Mini-Drop Administration Sets With Check Valve

More than 10 numbers, contact manufacturer.

V1448

Primay ADDitIV® Mini-Drop Administration Sets With Check Valve

More than 10 numbers, contact manufacturer.

NF1290

Infusomat® Space® Pump Y-Type Blood IV Set With Anti-Siphon Valve

More than 10 numbers, contact manufacturer.

363419

Infusomat® Space® Pump Burette IV Set With Anti-Siphon Valve

More than 10 numbers, contact manufacturer.

363422

Primary IV Set

More than 10 numbers, contact manufacturer.

354205

ADDitIV® Administration Set

More than 10 numbers, contact manufacturer.

354206

Infusomat® Space® Pump IV Set With Anti-Siphon Valve

More than 10 numbers, contact manufacturer.

363433

Infusomat® Space® Pump IV Set With Anti-Siphon Valve

More than 10 numbers, contact manufacturer.

363420

Infusomat® Space® Pump IV Set With Anti-Siphon Valve

More than 10 numbers, contact manufacturer.

363424

Infusomat® Space® Pump IV Set With Anti-Siphon Valve

More than 10 numbers, contact manufacturer.

480256

Infusomat® Space® Pump IV Set With Anti-Siphon Valve

More than 10 numbers, contact manufacturer.

363421

Infusomat® Space® Pump IV Set With Anti-Siphon Valve

More than 10 numbers, contact manufacturer.

363423

Infusomat® Space® Pump IV Set With Anti-Siphon Valve

More than 10 numbers, contact manufacturer.

363430

SafeDAY ™ IV Administration Set

More than 10 numbers, contact manufacturer.

352636

SafeDAY ™ IV Administration Set

More than 10 numbers, contact manufacturer.

352640

SafeDAY ™ IV Administration Set

More than 10 numbers, contact manufacturer.

352635

SafeDAY ™ IV Administration Set

More than 10 numbers, contact manufacturer.

352639

SafeDAY ™ IV Administration Set

More than 10 numbers, contact manufacturer.

352643

Issue

The potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies

B. Braun Medical Inc.

824 12th Avenue, Bethlehem, Pennsylvania, United States, 18018

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Recall date
Identification number
RA-78355

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe