Health product recall

Azurion 7 M20

Last updated

Summary

Product
Azurion 7 M20
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products Lot or serial number Model or catalogue number
Azurion 7 M20 Not applicable. 722 224 722 079

Issue

Philips has identified that grease leaked from the bearings and/or anti-corrosion oil applied on the bearings can result in excessive lubrication on the ceiling rail. This can reduce the friction between the friction wheel and the rails. The reduced traction force causes the friction wheel to slip on the rail, leading to a "position slip" error mode being detected by the software. When this error occurs, the x-y-rz joystick is unable to send motorized requests to the relevant axes. Due to this issue, the motorized longitudinal movement of the Flexarm stand may be inconsistent (not smooth) and eventually may become unavailable.

Recall start date: August 7, 2024

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Radiology
Companies
Philips Electronics Ltd.
1875 Buckhorn Gate, Mississauga, Ontario, Canada, L4W 5P1
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-75967

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe