AXIOM Vertix and AXIOM Multix
Brand(s)
Last updated
Summary
Product
AXIOM Vertix and AXIOM Multix
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
| Affected products | Lot or serial number | Model or catalog number |
|---|---|---|
| AXIOM Vertix MD Trauma | 8908290 | 8908290 |
| AXIOM Multix Mt | 8395381 8395415 8395399 | 44 99 302 |
Issue
Siemens Healthineers was informed by the monitor supplier that 5 out of 27.000 delivered power supplies of a touch-display power supply unit were affected by overheating due to a short circuit. In one of the five events, the connected monitor caught fire. No injuries were associated with these events. To ensure the continued safe operation of your system and to safeguard its lifetime, the power supply of the touch display will be replaced preventively.
Recall start date: May 29, 2024
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Radiology
Companies
Siemens AG
Wittelsbacherplatz 2, Muenchen, Germany, 80333
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-75669
Get notified