Health product recall

AVVIGO™+ Multi-Modality Guidance System

Last updated

Summary

Product
AVVIGO™+ Multi-Modality Guidance System
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products Lot or serial number Model or catalog number
AVVIGO™ + Mobile System All lots. H7492493120C0
AVVIGO™ + Integrated System All lots. H7492493120I0
AAVVIGO™ + Mobile Evaluation System All lots. H7492493120D0

Issue

On November 25th, 2024, BSC notified users of a potential software anomaly that can occur when the Avvigo+ Multi-Modality Guidance System is connected to an OptiCross 18 Imaging Catheter. When in live or record mode, the grid mark overlay on cross-sectional view would not display correctly. This advisory is to inform customers of a software update to correct this anomaly.

Recall start date: November 18, 2024

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Radiology
Companies

Boston Scientific Corporation

300 Boston Scientific Way, Marlborough, Massachusetts, United States, 01752

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-76712