AVEA Ventilator (2015-04-28)

Starting date:
April 28, 2015
Posting date:
June 15, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-53805

Recalled Products

  1. AVEA Ventilator

Reason

Avea ventilators can, over-time, develop a pressure transducer failure where the ventilator activates a false extended high peak or circuit occlusion alarm. This will cause safety value to immediately open and stop device from ventilating. (193 devices/154 spare parts affected)

Affected products

A. AVEA Ventilator

Lot or serial number

All lots.

Model or catalog number
  • AVEA
Companies
Manufacturer

Carefusion

22745 Savi Ranch Parkway

Yorba Linda

92887

California

UNITED STATES