This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

AVEA Ventilator (2015-04-28)

Starting date:
April 28, 2015
Posting date:
June 10, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-53735

Recalled Products

A. AVEA Ventilator

Reason

AVEA ventilators can, over-time, develop a pressure transducer failure where the ventilator activates a false extended high peak or circuit occlusion alarm. This will cause safety value to immediately open and stop device from ventilating. (193 devices/154 spare parts affected)

Affected products

A. AVEA Ventilator

Lot or serial number

All lots.

Model or catalog number

AVEA

Companies
Manufacturer

Carefusion

22745 Savi Ranch Parkway

Yorba Linda

92887

California

UNITED STATES