Health product recall

Avance and Aespire Anethiesia Machines

Last updated

Summary

Product
Avance and Aespire Anethiesia Machines
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information.

Audience
Healthcare

Affected products

Affected products

Lot or Serial number

Model or Catalogue number

Avance Cs2 Anesthesia Machine

More than 10 numbers, contact manufacturer.

1009-9050-000

Avance Anesthesia System

More than 10 numbers, contact manufacturer.

1009-9002-000

Aespire View Anesthesia Machine

More than 10 numbers, contact manufacturer.

1009-9212-000

Issue

Avance Cs2, Avance Cs2 pro anesthesia devices and field replacement batteries for Avance Cs2, Avance Cs2 pro, Avance, Amingo, Aespire View - a potential battery issue can result in premature shutdown of the anesthesia device in situations where ac mains power is lost and back up emergency power is not available.

Recall start date: May 19, 2022

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Anaesthesiology
Companies

Datex-Ohmeda, Inc.

3030 Ohmeda Drive, Madison, Wisconsin, United States, 53707-7550

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-64207

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe