Automated Impella Controller
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
Automated Impella Controller |
Software version 8.2.3 |
0042-0040-CA |
Automated Impella Controller |
Software version 8.2.3 |
0042-0000-CA |
Issue
Abiomed has identified that Automated Impella Controller (AICS) with software V8.2.3, when used in conjunction with an Impella 5.5 with smart assist set (product code 0550-0004), has a delayed "purge system blocked" alarm. Software versions post-v10.1 have this issue addressed.
Abiomed recommends that you do not use the AIC with an Impella 5.5 with smart assist set (product code 0550-0004) unless AIC software is updated to v10.1 or above.
Additional information
Details
Abiomed Inc.
22 Cherry Hill Drive, Danvers, Massachusetts, United States, 01923