Health product recall

AutoMate 2500 Family (2018-01-12)

Starting date:
January 12, 2018
Posting date:
February 9, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-65930

Affected products

AUTOMATE 2500 FAMILY

Reason

Belts on the x-axis (and on the y-axis, to a lesser extent) manufactured from 2015 are showing premature failure, posing the remote risk of cross-contamination.

Affected products

AUTOMATE 2500 FAMILY

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number
  • ODL25120
  • ODL25125
  • ODL25250
  • ODL25255
Companies
Manufacturer
Beckman Coulter, Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES