Autoimmune EIA Anti-Tissue Transglutaminase (tTG) IgA (2019-04-10)
- Starting date:
- April 10, 2019
- Posting date:
- May 10, 2019
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-69866
Last updated:
Affected Products
Autoimmune EIA Anti-Tissue Transglutaminase (tTG) IgA
Reason
Elevated lot-to-lot bias were observed by comparing a new EIA anti-Q15tTG IgA autoimmune lot (catalog# 4252260, lot# 18200a) to a previous lot (catalog# 4252260, lot# 18010a). This observation has been confirmed by recent tests by Bio-Rad using internal test samples.
Affected products
Autoimmune EIA Anti-Tissue Transglutaminase (tTG) IgA
Lot or serial number
18200A
Model or catalog number
4252260
Companies
- Manufacturer
-
Bio-Rad Laboratories, Inc.
4000 Alfred Nobel Dr.
Hercules
94547
California
UNITED STATES