AU5800 Clinical Chemistry Analyzer
- Starting date:
- November 26, 2012
- Posting date:
- December 24, 2012
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-25919
Recalled Products
- AU5800 Clinical Chemistry Analyzer
Reason
Under certain conditions, measurements of the Reagent Blank (RB) or Calibration (Cal) sample results are not saved and the software reverts to a previously stored, acceptable RB/Cal.
Affected products
A. AU5800 Clinical Chemistry Analyzer
Lot or serial number
All Lots
Model or catalog number
AU5810, AU5820, AU5830, AU5840