AU5800 Clinical Chemistry Analyzer
- Starting date:
- October 29, 2012
- Posting date:
- November 26, 2012
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public (GP), Healthcare Professionals, Hospitals
- Identification number:
- RA-15908
Recalled Products
- AU5800 Clinical Chemistry Analyzer
Reason
If the AU5800 diluted detergent tank valve is positioned to "closed" during the instrument "measure" mode, diluted detergent will not be dispensed and may adversely affect washing of cuvettes, probes and mixing-bars.
Affected products
A. AU5800 Clinical Chemistry Analyzer
Lot or serial number
All Lots
Model or catalog number
AU5810
AU5820
AU5830
AU5840
Companies
- Manufacturer
- Beckman Coulter Inc.