AU5400 Chemistry Analyzer
- Starting date:
- May 14, 2012
- Posting date:
- June 11, 2012
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-26315
Recalled Products
- AU5400 Chemistry Analyzer
Reason
Although the device will indicate when the Ion Selective Electorde (ISE) Sample Probe is fully occluded or near fully occluded as a result of a clot, it may not alarm when only partially occluded.
Affected products
A. AU5400 Chemistry Analyzer
Lot or serial number
All Lots
Model or catalog number
- AU5400
- AU5421
- AU5431
Companies
- Manufacturer
- Beckman Coulter, Inc.