This page has been archived on the Web
Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.
AU5400 Chemistry Analyzer
- Starting date:
- May 14, 2012
- Posting date:
- June 11, 2012
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-26315
Recalled Products
- AU5400 Chemistry Analyzer
Reason
Although the device will indicate when the Ion Selective Electorde (ISE) Sample Probe is fully occluded or near fully occluded as a result of a clot, it may not alarm when only partially occluded.
Affected products
A. AU5400 Chemistry Analyzer
Lot or serial number
All Lots
Model or catalog number
- AU5400
- AU5421
- AU5431
Companies
- Manufacturer
-
Beckman Coulter, Inc.