Attune System - Tibial Insert Fixed Bearing Posterior Stabilized
Brand(s)
Last updated
Summary
Product
Attune System - Tibial Insert Fixed Bearing Posterior Stabilized
Issue
Medical devices - Performance
What to do
Contact manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
---|---|---|
Attune System - Tibial Insert Fixed Bearing Posterior Stabilized |
JN6613 |
1516-40-507 |
Issue
The subject lot was sent for sterilization twice. This exceeds the validated range for exposure to gamma radiation of these devices and may result in changes to the tibial insert material properties.
Recall start date: February 23, 2023
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Orthopaedics
Companies
Depuy (Ireland)
Loughbeg, Ringaskiddy, Cork, Ireland
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-73008
Get notified
Receive notifications for new and updated recalls and alerts by category.