Atlan® A350
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number/ |
---|---|---|
Atlan® A350 XL |
ASRJ-0212 |
8621600 |
Atlan® A350 |
More than 10 numbers, contact manufacturer. |
8621500 |
Issue
The recall is being initiated as the market surveillance activities revealed a few cases in which the internal backup battery failed spontaneously while the Atlan® is being operated without mains supply. This resulted in an unexpected shutdown of the device while it was running on batteries. The affected Atlans started with a battery status of 100%, but quickly shut down and did not trigger the alarm for low battery. A secondary acoustic alarm signal which is independent from mains and battery power was generated as specified. No patient consequences have been reported to Draeger so far.
Recall start date: November 21, 2023
Additional information
Details
Draegerwerk AG & Co. KGaA
Moislinger Allee 53-55, Lubeck, Germany, 23542