Health product recall

Atellica UAS 800 Analyzer (2019-09-13)

Starting date:
September 13, 2019
Posting date:
October 11, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-71273


Last updated:

Affected Products

Atellica UAS 800 Analyzer

Reason

Siemens Healthcare Diagnostics has confirmed that a communication error could be generated on the system application software when using the ASTM interface protocol to connect with laboratory information system (LIS). The following error messages, "LIS connection lost (412)" or "LIS communication error (224)," may potentially be generated when the host query response or the worklist entry message from the LIS and the result message from the system application software occur simultaneously.

Affected products

Atellica UAS 800 Analyzer

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number
  • 11065004
Companies
Manufacturer

Siemens Healthcare Diagnostics Inc.

511 Benedict Ave

Tarrytown

10591

New York

UNITED STATES