Atellica CH Reaction Ring Cuvette Segment
Brand(s)
Last updated
Summary
Product
Atellica CH Reaction Ring Cuvette Segment
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Affected products
Affected products | Lot or serial number | Model or catalog number |
---|---|---|
Atellica CH Reaction Ring Cuvette Segment | Starting with “N15” or “15" | 11099326 |
Issue
Siemens Healthineers has confirmed, through the internal investigation of customer complaints that there is a potential for imprecision with seven (7) assays identified to have optical sensitivity when using a subset of lots of Atellica® CH Reaction Ring Cuvette segments on some Atellica CH analyzers.
Recall Start Date: October 28, 2024
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Ophthalmology
Companies
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave, Tarrytown, New York, United States, 10591
Published by
Health Canada
Audience
General public
Healthcare
Industry
Recall class
Type II
Identification number
RA-76440
Get notified