Health product recall

Atellica CH 930 Analyzer (2019-09-25)

Starting date:
September 25, 2019
Posting date:
October 11, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-71265



Last updated: 2019-10-11

Affected Products

Atellica CH 930 Analyzer

Reason

Siemens Healthcare Diagnostics, INC. has confirmed that on the Atellica CH 930 Analyzer, when a reagent lot calibration and a pack calibration (C0 adjust) on the same assay are manually ordered at the same time, the reagent lot calibration recovery of the analyte is 5 times higher because the 5x predilution factor is improperly factored into the QC and patient sample result calculation. The following assays are affected: acetaminophen (ACET), cholesterol (CHOL_2), glucose oxidase (GLUO), triglycerides (TRIG), uric acid (UA), urea nitrogen (UN_C) and inorganic phosphorus (IP). If QC is ordered with calibration and C0 adjust, QC results will be out of range high (red status on the atellica solution - calibration results details screen) and the calibration will display a status of "Awaiting Acceptance". If QC is not ordered with calibration and C0 adjust, then the calibration status will be shown as "Valid". Patient results and QC results will be 5 times higher.

Affected products

Atellica CH 930 Analyzer

Lot or serial number
  • CM00237
  • CM00238
  • CM00239
  • CM00416
  • CM01054
  • CM01112
  • CM01178

     

Model or catalog number
  • 11067000
Companies
Manufacturer

Siemens Healthcare Diagnostics Inc.

511 Benedict Ave

Tarrytown

10591

New York

UNITED STATES