Atellica CH 930 Analyzer (2019-09-25)
- Starting date:
- September 25, 2019
- Posting date:
- October 11, 2019
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-71265
Last updated: 2019-10-11
Affected Products
Atellica CH 930 Analyzer
Reason
Siemens Healthcare Diagnostics, INC. has confirmed that on the Atellica CH 930 Analyzer, when a reagent lot calibration and a pack calibration (C0 adjust) on the same assay are manually ordered at the same time, the reagent lot calibration recovery of the analyte is 5 times higher because the 5x predilution factor is improperly factored into the QC and patient sample result calculation. The following assays are affected: acetaminophen (ACET), cholesterol (CHOL_2), glucose oxidase (GLUO), triglycerides (TRIG), uric acid (UA), urea nitrogen (UN_C) and inorganic phosphorus (IP). If QC is ordered with calibration and C0 adjust, QC results will be out of range high (red status on the atellica solution - calibration results details screen) and the calibration will display a status of "Awaiting Acceptance". If QC is not ordered with calibration and C0 adjust, then the calibration status will be shown as "Valid". Patient results and QC results will be 5 times higher.
Affected products
Atellica CH 930 Analyzer
Lot or serial number
- CM00237
- CM00238
- CM00239
- CM00416
- CM01054
- CM01112
-
CM01178
Model or catalog number
- 11067000
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
10591
New York
UNITED STATES