Astral 150™
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
Astral 150™ |
More than 10 numbers, contact manufacturer. |
27073 |
Astral 150™ |
More than 10 numbers, contact manufacturer. |
27003 |
Issue
An internal electrical component (supercapacitor) may leak electrolyte over time. In rare cases, this may damage specific circuitry on the ventilator's Printed Circuit Board Assembly ("PCBA"), causing the ventilator to inadvertently enter a fail-safe state. In this state, the ventilator is no longer able to deliver therapy. Alternative means of ventilation must be provided.
Additional information
Details
Resmed Pty Limited
1 Elizabeth Macarthur Drive, Bella Vista, New South Wales, Australia, 2153