Summary

Product
Astral 150™
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information.

Affected products

Affected products Lot or serial number Model or catalogue number
Astral 150™ More than 10 numbers, contact manufacturer. 27073
Astral 150™ More than 10 numbers, contact manufacturer. 27003

Issue

An internal electrical component (supercapacitor) may leak electrolyte over time. In rare cases, this may damage specific circuitry on the ventilator's Printed Circuit Board Assembly ("PCBA"), causing the ventilator to inadvertently enter a fail-safe state. In this state, the ventilator is no longer able to deliver therapy. Alternative means of ventilation must be provided.

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Anaesthesiology
Companies

Resmed Pty Limited

1 Elizabeth Macarthur Drive, Bella Vista, New South Wales, Australia, 2153

Published by
Health Canada
Audience
General public
Healthcare
Recall class
Type II
Recall date
Identification number
RA-82300