Astral 150™
Brand(s)
Last updated
Summary
Product
Astral 150™
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Affected products
| Affected products | Lot or serial number | Model or catalogue number |
|---|---|---|
| Astral 150™ | More than 10 numbers, contact manufacturer. | 27073 |
| Astral 150™ | More than 10 numbers, contact manufacturer. | 27003 |
Issue
An internal electrical component (supercapacitor) may leak electrolyte over time. In rare cases, this may damage specific circuitry on the ventilator's Printed Circuit Board Assembly ("PCBA"), causing the ventilator to inadvertently enter a fail-safe state. In this state, the ventilator is no longer able to deliver therapy. Alternative means of ventilation must be provided.
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Anaesthesiology
Companies
Resmed Pty Limited
1 Elizabeth Macarthur Drive, Bella Vista, New South Wales, Australia, 2153
Published by
Health Canada
Audience
General public
Healthcare
Recall class
Type II
Recall date
Identification number
RA-82300