Health product recall

Astra, Azure (XT SR, XT DR, S SR, S DR) MRI SureScan, Percepta, Serena, Solara (Quad CRT-P, CRT-P) MRI Sure Scan (2019-05-06)

Starting date:
May 6, 2019
Posting date:
May 24, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-70025


Last updated:

Affected Products

  1. Astra XT SR MRI SureScan
  2. Astra XT DR MRI SureScan
  3. Astra S SR MRI SureScan
  4. Astra S DR MRI SureScan
  5. Azure XT SR MRI SureScan
  6. Azure XT DR MRI SureScan
  7. Azure S DR MRI SureScan
  8. Azure S SR MRI SureScan
  9. Percepta CRT-P MRI SureScan
  10. Percepta Quad CRT-P MRI SureScan
  11. Serena CRT-P MRI SureScan
  12. Serena Quad CRT-P MRI SureScan
  13. Solara CRT-P MRI SureScan
  14. Solara Quad CRT-P MRI SureScan

Reason

Medtronic has identified a rare but potentially serious failure mode in a population of Azure and Astra pacemakers, and Percepta, Serena and Solara cardiac resynchronization therapy pacemakers (CRT-P), manufactured with a specific multilayer ceramic capacitor. As of April 26, 2019, three complaints out of ~266,700 devices distributed worldwide since February 2017, have been received that included a no output /no telemetry scenario resulting from rapid battery depletion. Battery depletion due to this issue can range from several days to several weeks. One of these reported events contributed to a patient death. The three confirmed failures occurred within 9 months post implant. The projected rate for this issue is 0.0028%, with the most susceptible period for a leakage pathway to develop in the capacitor being the first 12 months post implant. Upon further analysis at the Medtronic Product Analysis Lab (PAL), it was determined that the device failures were both caused by a cracked XC342 (C12) capacitor. If a crack in the ceramic capacitor, along with humidity and bias, results in silver migration that creates a leakage path, the battery is at risk of becoming depleted without warning to the patient.

Affected products

A. Astra XT SR MRI SureScan

Lot or serial number

Not applicable.

Model or catalog number

X2SR01

Companies
Manufacturer

Medtronic Inc.

710 Medtronic Parkway

Minneapolis

55432

Minnesota

UNITED STATES



B. Astra XT DR MRI SureScan

Lot or serial number

Not applicable.

Model or catalog number

X2DR01

Companies
Manufacturer

Medtronic Inc.

710 Medtronic Parkway

Minneapolis

55432

Minnesota

UNITED STATES



C. Astra S SR MRI SureScan

Lot or serial number

Not applicable.

Model or catalog number

X3SR01

Companies
Manufacturer

Medtronic Inc.

710 Medtronic Parkway

Minneapolis

55432

Minnesota

UNITED STATES



D. Astra S DR MRI SureScan

Lot or serial number

Not applicable.

Model or catalog number

X3DR01

Companies
Manufacturer

Medtronic Inc.

710 Medtronic Parkway

Minneapolis

55432

Minnesota

UNITED STATES



E. Azure XT SR MRI SureScan

Lot or serial number

Not applicable.

Model or catalog number

W2SR01

Companies
Manufacturer

Medtronic Inc.

710 Medtronic Parkway

Minneapolis

55432

Minnesota

UNITED STATES



F. Azure XT DR MRI SureScan

Lot or serial number

Not applicable.

Model or catalog number

W2DR01

Companies
Manufacturer

Medtronic Inc.

710 Medtronic Parkway

Minneapolis

55432

Minnesota

UNITED STATES



G. Azure S DR MRI SureScan

Lot or serial number

Not applicable.

Model or catalog number

W3DR01

Companies
Manufacturer

Medtronic Inc.

710 Medtronic Parkway

Minneapolis

55432

Minnesota

UNITED STATES



H. Azure S SR MRI SureScan

Lot or serial number

Not applicable.

Model or catalog number

W3SR01

Companies
Manufacturer

Medtronic Inc.

710 Medtronic Parkway

Minneapolis

55432

Minnesota

UNITED STATES



I. Percepta CRT-P MRI SureScan

Lot or serial number

Not applicable.

Model or catalog number

W1TR04

Companies
Manufacturer

Medtronic Inc.

710 Medtronic Parkway

Minneapolis

55432

Minnesota

UNITED STATES



J. Percepta Quad CRT-P MRI SureScan

Lot or serial number

Not applicable.

Model or catalog number

W4TR04

Companies
Manufacturer

Medtronic Inc.

710 Medtronic Parkway

Minneapolis

55432

Minnesota

UNITED STATES



K. Serena CRT-P MRI SureScan

Lot or serial number

Not applicable.

Model or catalog number

W1TR05

Companies
Manufacturer

Medtronic Inc.

710 Medtronic Parkway

Minneapolis

55432

Minnesota

UNITED STATES



L. Serena Quad CRT-P MRI SureScan

Lot or serial number

Not applicable.

Model or catalog number

W4TR05

Companies
Manufacturer

Medtronic Inc.

710 Medtronic Parkway

Minneapolis

55432

Minnesota

UNITED STATES



M. Solara CRT-P MRI SureScan

Lot or serial number

Not applicable.

Model or catalog number

W1TR06

Companies
Manufacturer

Medtronic Inc.

710 Medtronic Parkway

Minneapolis

55432

Minnesota

UNITED STATES



N. Solara Quad CRT-P MRI SureScan

Lot or serial number

Not applicable.

Model or catalog number

W4TR06

Companies
Manufacturer

Medtronic Inc.

710 Medtronic Parkway

Minneapolis

55432

Minnesota

UNITED STATES