Health product recall

Asserachrom® HPIA

Last updated

Summary

Product
Asserachrom® HPIA
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products Lot or serial number Model or catalogue number
Asserachrom® HPIA 271288 00615

Issue

After receiving three customer complaints (from outside Canada and the US), Stago investigated and confirmed that approximately 5% of reagent 1 strips had a defect for Asserachrom® HPIA  lot 271288.  When the defect is present, every well in the impacted strip produces negative results.  Kit reference reagent 6 and control 2 (reagent 7b) produce low optical density (OD) readings, similar to those of control 1 (reagent 7a).  As a result, these values will not meet the requirements specified in the assigned value sheet for the lot and must not be used. 
consignee actions:

  • Stop using lot 271288 and dispose of all remaining product according to local regulations. 
  • Complete, sign, and return acknowledgement form. 

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Immunology
Companies

Diagnostica Stago SAS

3 Allee Theresa, Asnieres Sur Seine, France, 92600

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Recall date
Identification number
RA-81397