Asserachrom® HPIA
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
| Affected products | Lot or serial number | Model or catalogue number |
|---|---|---|
| Asserachrom® HPIA | 271288 | 00615 |
Issue
After receiving three customer complaints (from outside Canada and the US), Stago investigated and confirmed that approximately 5% of reagent 1 strips had a defect for Asserachrom® HPIA lot 271288. When the defect is present, every well in the impacted strip produces negative results. Kit reference reagent 6 and control 2 (reagent 7b) produce low optical density (OD) readings, similar to those of control 1 (reagent 7a). As a result, these values will not meet the requirements specified in the assigned value sheet for the lot and must not be used.
consignee actions:
- Stop using lot 271288 and dispose of all remaining product according to local regulations.
- Complete, sign, and return acknowledgement form.