Health product recall

ASSAYED CHEMISTRY CONTROL PREMIUM PLUS (2019-07-23)

Starting date:
July 23, 2019
Posting date:
August 2, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-70611


Last updated:

Affected Products

ASSAYED CHEMISTRY CONTROL PREMIUM PLUS

Reason

Randox have realigned the RX Analyser Series calibration targets by 10% for AST and 9% for ALT in line with the Mean of All Instrument method target from calibrator lot 961UE onwards. No other instruments are affected. The target and range for the associated Quality Control Material has been revised to align with this change. Updated RX control targets are available at randox.com.

Affected products

ASSAYED CHEMISTRY CONTROL PREMIUM PLUS

Lot or serial number

961UE

995UE

Model or catalog number

CAL2351

Companies
Manufacturer

Randox Laboratories Ltd.

55 Diamond Road

Crumlin, County Antrim

BT29 4QY

UNITED KINGDOM