This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Asena GH and Alaris Syringe Pumps (2017-05-04)

Starting date:
May 4, 2017
Posting date:
June 5, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
Healthcare Professionals, General Public, Hospitals
Identification number:
RA-63490

Affected Products

A. Asena GH

B. Alaris PK Syringe Pump (Rebranding)

C. Alaris GH Syringe Pump With Guardrails Safety Software (Rebranding)

D. Alaris CC Syringe Pump With Guardrails Safety Software (Rebranding)

E. Alaris CC Syringe Pump (Rebranding)

F. Alaris Tiva Syringe Pump (Rebranding)

G. Alaris GH Syringe Pump (Rebranding)

Reason

While the pump is infusing, a breakage of the plunger backplate spring may allow movement of the syringe plunger that may result in siphonage. Neonatal and paediatric patients, or those receiving critical drugs at low infusion rates, would be considered to be the most at risk if small volumes of fluids reach the patient due to siphoning.

Affected products

A. Asena GH

Lot or serial number

More than 10 numbers, contact manufacturer

Model or catalog number

80023FR01

80023GB01

Companies
Manufacturer

Carefusion Switzerland 317 Sarl

A-One Business Centre, Zone D'Activites, Vers-La-Piece No 10

Rolle

CH-1180

SWITZERLAND



B. Alaris PK Syringe Pump (Rebranding)

Lot or serial number

More than 10 numbers, contact manufacturer

Model or catalog number

80053UN01

8005PK201

8005TIG01

Companies
Manufacturer

Carefusion Switzerland 317 Sarl

A-One Business Centre, Zone D'Activites, Vers-La-Piece No 10

Rolle

CH-1180

SWITZERLAND



C. Alaris GH Syringe Pump With Guardrails Safety Software (Rebranding)

Lot or serial number

More than 10 numbers, contact manufacturer

Model or catalog number

8002MED01-G

Companies
Manufacturer

Carefusion Switzerland 317 Sarl

A-One Business Centre, Zone D'Activites, Vers-La-Piece No 10

Rolle

CH-1180

SWITZERLAND



D. Alaris CC Syringe Pump With Guardrails Safety Software (Rebranding)

Lot or serial number

More than 10 numbers, contact manufacturer

Model or catalog number

80033UND1-G

Companies
Manufacturer

Carefusion Switzerland 317 Sarl

A-One Business Centre, Zone D'Activites, Vers-La-Piece No 10

Rolle

CH-1180

SWITZERLAND



E. Alaris Cc Syringe Pump (Rebranding)

Lot or serial number

More than 10 numbers, contact manufacturer

Model or catalog number

80033UND1

Companies
Manufacturer

Carefusion Switzerland 317 Sarl

A-One Business Centre, Zone D'Activites, Vers-La-Piece No 10

Rolle

CH-1180

SWITZERLAND



F. Alaris Tiva Syringe Pump (Rebranding)

Lot or serial number

More than 10 numbers, contact manufacturer

Model or catalog number

80043UN01

Companies
Manufacturer

Carefusion Switzerland 317 Sarl

A-One Business Centre, Zone D'Activites, Vers-La-Piece No 10

Rolle

CH-1180

SWITZERLAND



G. Alaris GH Syringe Pump (Rebranding)

Lot or serial number

More than 10 numbers, contact manufacturer

Model or catalog number

80023UN01

Companies
Manufacturer

Carefusion Switzerland 317 Sarl

A-One Business Centre, Zone D'Activites, Vers-La-Piece No 10

Rolle

CH-1180

SWITZERLAND