Health product recall

Artis Zeego System (2020-02-17)

Starting date:
February 17, 2020
Posting date:
October 9, 2020
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-74085

Last updated: 2020-10-09

Affected Products

A. ARTIS ZEEGO MULTI-AXIS

B. ARTIS ZEE FLOOR

C. ARTIS ZEE CEILING

D. ARTIS ZEE BIPLANE

E. ARTIS Q

F. ARTIS Q.ZEN MODULAR ANGIOGRAPHIC SYSTEM

Reason

Siemens has identified 3 software issues during regular field observations that could lead to the Artis zeego systems to block all stand and/or table movements until a technician reactivates them.

Affected products

A. ARTIS ZEEGO MULTI-AXIS

Lot or serial number

160846
161205

Model or catalog number

10280959

Companies
Manufacturer
Siemens Healthcare GMBH
Henkestr. 127
Erlangen
91052
GERMANY

B. ARTIS ZEE FLOOR

Lot or serial number

124222
136972
137423

Model or catalog number

10094135

Companies
Manufacturer
Siemens Healthcare GMBH
Henkestr. 127
Erlangen
91052
GERMANY

C. ARTIS ZEE CEILING

Lot or serial number

147638
147832
147908
148012
148024
148204
148244
148304

Model or catalog number

10094137

Companies
Manufacturer
Siemens Healthcare GMBH
Henkestr. 127
Erlangen
91052
GERMANY

D. ARTIS ZEE BIPLANE

Lot or serial number

153810
154679

Model or catalog number

10094141

Companies
Manufacturer
Siemens Healthcare GMBH
Henkestr. 127
Erlangen
91052
GERMANY

E. ARTIS Q

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

10848280
10848281
10848282
10848283

Companies
Manufacturer
Siemens Healthcare GMBH
Henkestr. 127
Erlangen
91052
GERMANY

F. ARTIS Q.ZEN MODULAR ANGIOGRAPHIC SYSTEM

Lot or serial number

123048
123214

Model or catalog number

10848355

Companies
Manufacturer
Siemens Healthcare GMBH
Henkestr. 127
Erlangen
91052
GERMANY