Artis Zeego Multi-Axis (2014-09-22)
- Starting date:
- September 22, 2014
- Posting date:
- September 26, 2014
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-41535
Recalled products
- Artis Zeego Multi-Axis
Reason
- A cable inside the C-arm of systems affected may break within the specific serial number range of 1000 - 1132. A cable break may cause restricted functionality or even failing of the system. An ongoing procedure may be terminated at this system.
- A cable inside the C-arm of systems within the serial range of 1000 - 1599, may have been routed in an inappropriate manner. Without additional measure, there is potential that a cable inside the C-arm may break. The problem will not occur systematically but sporadically in the ongoing life time of a system.
Affected products
A. Artis Zeego Multi-Axis
Lot or serial number
- S/N 160109
- S/N 160354
- S/N 160423
- S/N 160810
- S/N 160846
Model or catalog number
- 10280959
Companies
- Manufacturer
-
Siemens AG
Wittelsbacherplatz 2
Muenchen
80333
GERMANY