This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Artis Zeego and Artis Zee Multi-purpose, Floor, Celing and Biplane (2014-06-12)

Starting date:
June 12, 2014
Posting date:
June 24, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-40185

Recalled products

  1. Artis Zeego
  2. Artis Zee Floor
  3. Artis Zee Ceiling
  4. Artis Zee Biplane

Reason

Siemens is informing all users of Artis Zee and Artis Zeego systems that under certain circumstances, the release of radiation can become blocked unnecessarily. This situation comes about after the system has booted up, and after an error with the large display occurs that causes the system to block the release of radiation. After this, the system remains in a state where radiation cannot be released, even if the error that triggered this state was only a temporary one, and the error condition itself goes away after a few seconds (such as a brief communication error).

Affected products

A. Artis Zeego

Lot or serial number
  • 160810
  • 160846
Model or catalog number
  • 10280959
Companies
Manufacturer

Siemens AG

Wittelsbacherplatz 2

Muenchen

80333

GERMANY


B. Artis Zee Floor

Lot or serial number
  • 136972
  • 137167
Model or catalog number
  • 10094135
Companies
Manufacturer

Siemens AG

Wittelsbacherplatz 2

Muenchen

80333

GERMANY


C. Artis Zee Ceiling

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number
  • 10094137
Companies
Manufacturer

Siemens AG

Wittelsbacherplatz 2

Muenchen, Germany

80333


D. Artis Zee Biplane

Lot or serial number
  • 153810
  • 153854
  • 153859
  • 154569
  • 154679
  • 154691
  • 154693
Model or catalog number
  • 10094141
Companies
Manufacturer

Siemens AG

Wittelsbacherplatz 2

Muenchen

80333

GERMANY