This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Artis Zee Floor (2017-02-06)

Starting date:
February 6, 2017
Posting date:
March 10, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-62448

Affected Products

  1. Artis Zee Floor Mn
  2. Artis Zee Floor

Reason

Due to a component defect on Artis Systems with an A100Plus Generator of a certain delivery lot and 2-Focus Megalix Cat Plus Tube Unit, a module in the high voltage generator can fail. If the fault occurs, result can be failure of large focus. Small focus will still be available for continued operation.

Affected products

  1. Artis Zee Floor Mn

Lot or serial number

All lots

Model or catalog number
  • 10094142
Companies
Manufacturer

Siemens Healthcare GMBH

HENKESTRASSE 127

ERLANGEN

91052

GERMANY



B. Artis Zee Floor

Lot or serial number

All lots

Model or catalog number

10094135

7412807

Companies
Manufacturer

Siemens Healthcare GMBH

HENKESTRASSE 127

ERLANGEN

91052

GERMANY