Health product recall

Artis zee Ceiling System (2018-10-03)

Starting date:
October 3, 2018
Posting date:
October 19, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-68046

Affected products

Artis zee Ceiling System

Reason

During regular product monitoring abrasion was identified at the cabling at some Artis Zee Ceiling Systems. Abrasion may occur in the area of the cable outlet at the interior C-Arm. This is not a systematic issue but occurs sporadically at single units during the lifetime of a system. Siemens will be exchanging the existing cable routing at the C-Arm to prevent cable damage in future.

Affected products

Artis zee Ceiling System

Lot or serial number

146106
146230
146314
146708

Model or catalog number

10094137