ARTIS ZEE and Q
Last updated
Summary
Product
ARTIS ZEE and Q
Issue
Medical devices - Software defect
What to do
Contact the manufacturer if you require additional information
Affected products
Affected Products |
Lot or serial number |
Model or catalog number |
---|---|---|
ARTIS ZEE MULTI-PURPOSE |
159022 |
10094139 |
ARTIS Q |
110070 |
10848281 |
Issue
If SID (source-to-image distance) lift movement is activated and X-ray shall be released simultaneously then X-ray is not possible and the message "no X-ray, try again" is displayed. X-ray is not possible. However, after stopping SID lift movement and releasing the X-ray switch, X-ray is possible again. Recall start date: 2021-11-03 |
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Radiology
Companies
SIEMENS HEALTHCARE LIMITED
1577 NORTH SERVICE ROAD EAST
OAKVILLE, ON CANADA
L6H 0H6
Published by
Health Canada
Audience
General public
Health professionals
Hospitals
Recall class
Type II
Identification number
RA-63634
Get notified