Artis Pheno System (2020-05-23)
- Starting date:
- May 23, 2020
- Posting date:
- May 23, 2020
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-73149
Last updated: 2020-05-22
Affected Products
Artis Pheno System
Reason
In the event of a fault within the drivetrain of an axis, the potential exists for the C-arm to leave its intended travel path by up to 10cm. The problem has not been reported at customer sites and is currently classified as a single event.
Affected products
Artis Pheno System
Lot or serial number
164101
164123
164221
Model or catalog number
Not applicable.
Companies
- Manufacturer
-
Siemens Healthcare GMBH
Henkestr. 127
Erlangen
91052
GERMANY