ARTIS pheno (2021-05-26)
- Starting date:
- May 26, 2021
- Posting date:
- June 17, 2021
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-75813
Last updated:
2021-06-17
Affected Products
ARTIS pheno
Reason
The Câarm might leave its intended travel path by up to 22 cm (9"), in cases of a fault within the drivetrain of an axis and while performing a stepped peripheral imaging (PERISTEPPING/PERIVISION).
Affected products
ARTIS pheno
Lot or serial number
164101
164123
164221
Model or catalog number
10849000
Companies
- Manufacturer
-
Siemens Healthcare GMBH
Henkestr. 127
Erlangen
91052
GERMANY