Health product recall

Artis Icono and Pheno Systems

Last updated

Summary

Product
Artis Icono and Pheno Systems
Issue
Medical devices - Device compatibility
What to do

Contact the manufacturer if you require additional information.

Affected products

Affected products

Lot or serial number

Model or catalog number

Artis Icono Biplane

180415
180324

11327600

Artis Pheno

164101

10849000

Artis Icono Floor

170369
170382

11327700

Issue

During transversal table movement, it could happen that the rolling bearing which is part of the linear guide rail and enables the transversal table movement may get dislocated. The rolling bearing may fall out of the linear guide rail and thus the mechanical connection between table top and table base may get lost.

Recall start date: December 19, 2022

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Radiology
Companies

Siemens Healthcare Gmbh

Henkestr. 127, Erlangen, Germany, 91052

Published by
Health Canada
Audience
General public
Healthcare
Industry
Recall class
Type II
Identification number
RA-72507

Get notified

Receive notifications for new and updated recalls and alerts by category.

Subscribe