This page has been archived on the Web
Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.
ARROW INTERNATIONAL VPS ACCESS KIT WITH TUOHY BORST COMPONENT(2015-08-11)
- Starting date:
- August 11, 2015
- Posting date:
- August 24, 2015
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-54792
Affected Products
A. ARROW INTERNATIONAL VPS ACCESS KIT WITH TUOHY BORST COMPONENT
Reason
Arrow is recalling certain lots of the products referenced above due to reports clinicians have had difficulty maintaining a secure connection between the male Luer of the Tuohy Borst Adapter of the vascular positioning system stylet to the hub of the catheter. When the connection is not secure, there is a possibility the connection will leak saline. This is most often discovered during initial flushing prior to insertion, in which case, a delay in treatment can occur while the health care provider locates a replacement device. In the unlikely event that the issue is discovered while the device is in use, it could result in an inability to obtain a signal on the VPS console. This would in turn necessitate removal and replacement of the catheter or an x-ray to confirm proper catheter placement.
Affected products
A. ARROW INTERNATIONAL VPS ACCESS KIT WITH TUOHY BORST COMPONENT
Lot or serial number
> 10 numbers, contact mfg
Model or catalog number
VPS7220B
Companies
- Manufacturer
-
Arrow International
Inc.
2400 Bernville Road
Pennsylvania
UNITED STATES