Argus Epic Camera (2013-10-31)
- Starting date:
- October 31, 2013
- Posting date:
- December 4, 2013
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-36987
Recalled products
A. Argus Epic Camera
Reason
A recent investigation has determined that visual inspections of the fork joints behind the detector that is conducted during six month preventive maintenance may not detect small cracks or deficiencies in the fork weldment.
Affected products
A. Argus Epic Camera
Lot or serial number
Not applicable
Model or catalog number
- 2145-3007A
Companies
- Manufacturer
-
Philips Medical Systems (Cleveland), Inc.
595 Miner Road
Cleveland
44143
Ohio
UNITED STATES