Health product recall

Arctic Sun 5000 Temperature Management System

Starting date:
August 25, 2017
Posting date:
September 11, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-64502

Reason

Bard is informing customers that use of the Arctic Sun Temperature Management System in excess of 2,000 system hours, without conducting preventative maintenance, may result in failure of certain components of the system, particularly in relation to the mixing pump, which runs continuously during operation. Failure of the mixing pump may cause the device to cease to function as intended, resulting in no availability of cooling or rewarming therapy for the patient.

Affected products

Arctic Sun 5000 Temperature Management System

Lot or serial number

More than 10 numbers, contact manufacturer

Model or catalog number

5000-00-00

Companies

Manufacturer

Medivance, Inc.

321 South Taylor avenue, Suite 200

Louisville

80027

Colorado

UNITED STATES