This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health professional risk communication

Archived – Important Safety Information Regarding the Use of SEVORANE AF (Sevoflurane) in Conjunction with Anesthesia Machines

Starting date:
November 17, 2003
Posting date:
January 8, 2004
Type of communication:
Dear Healthcare Professional Letter
Subcategory:
Drugs, Medical Device
Source of recall:
Health Canada
Issue:
Important Safety Information
Audience:
Healthcare Professionals
Identification number:
RA-17000820

This is duplicated text of a letter from Abbott Laboratories, Limited.

Contact the company for a copy of any references, attachments or enclosures.

Notice about Health Canada advisories

IMPORTANT SAFETY INFORMATION REGARDING THE USE OF SEVORANE AF (SEVOFLURANE) IN CONJUNCTION WITH ANESTHESIA MACHINES

November 17, 2003

Dear Healthcare Professional,

Abbott Laboratories, Limited ("Abbott") would like to make you aware of rare, isolated reports of smoke/fire or extreme heat which have been observed in anesthesia machines where SEVORANE AF* (sevoflurane) is used in conjunction with a desiccated carbon dioxide (CO2) absorbent. Although ongoing investigations have not yet identified the exact etiology, Abbott considers it important to inform anesthesia providers of the existence of these rare occurrences, which can result in patient injury.

Based on available information, it is important to highlight that in most cases reported to Abbott, desiccated CO2 absorbents were in use. It is also well documented that the exothermic reaction that occurs between all inhalational agents, including sevoflurane, and the CO2 absorber is increased when the absorber becomes desiccated, such as after an extended period of dry gas flow through the absorbent canister.

Please remember to continue to follow your best practices with respect to maintaining your anesthesia machines, such as:

  • If you suspect that the CO2 absorbent may be desiccated because it has not been in use for an extended period of time, replace it.
  • Although the exact conditions under which absorbent may become desiccated are not well defined, a low fresh gas flow rate over an extended period of non-use may contribute to unexpected desiccation of CO2 absorbent materials on the anesthesia machine. Therefore, completely shut off the anesthesia machine at the end of clinical use or after any case when a subsequent extended period of non-use is anticipated.
  • Turn off all vaporizers when not in use.
  • Verify the integrity of the packaging of new CO2 absorbents prior to use.
  • Periodically monitor the temperature of the CO2 absorbent canisters.
  • Monitor the correlation between the sevoflurane vaporizer setting and the inspired sevoflurane concentration. An unusually delayed rise or unexpected decline of inspired sevoflurane concentration compared to the vaporizer setting may be associated with excessive heating of the CO2 absorbent canister.

In addition, please consider the following:

  • The colour indicator of most CO2 absorbents does not necessarily change as a result of desiccation. A lack of significant colour change should therefore not be taken as an assurance of adequate hydration. CO2 absorbents should be replaced routinely regardless of the state of the colour indicator.
  • If excessive heat from the CO2 absorbent canister is noted, the clinical situation should be evaluated and disconnecting the patient from the anesthesia circuit should be considered.
  • The following findings have been reported in association with the rare events of fire and extreme heat:
    • Failed inhalation induction or inadequate anesthesia with sevoflurane
    • Patient signs of airway irritation, such as coughing
    • Oxygen desaturation, increased airway pressures, and difficult ventilation
    • Severe airway edema and erythema
    • Elevated carboxyhemoglobin levels
  • Current information indicates that typically these cases of fire or extreme heat were the first case of the day for the specific anesthesia machine, and the reported cases indicated that Baralyme® CO2 absorbent was used. However, cases of extreme heat have been reported in association with desiccated soda lime.
  • When desiccated CO2 absorbents are used with sevoflurane under extreme experimental conditions, flammable degradation products, including formaldehyde and methanol, may be present even in the absence of fire. The potential risk to patients receiving sevoflurane anesthesia due to these breakdown products has not been evaluated.1Footnote Similar to other potent inhalational anesthetic agents, sevoflurane, when exposed to desiccated CO2 absorbents, produces carbon monoxide.Footnote 2

SEVORANE AF* (sevoflurane) has been available in Canada since its initial launch in 1995. Although these events are rare and isolated, they can result in patient injury. Abbott will continue to investigate the causative and preventive factors surrounding the issues of fire, extreme heat, and potential breakdown products associated with the use of sevoflurane and desiccated CO2 absorbents. Abbott will continue to provide timely guidance as relevant information becomes available.

Please review the steps to reduce the risk of fire and extreme heat included in this letter with your anesthesia associates. As with all medical products, healthcare professionals are strongly encouraged to report any adverse events that occur with the use of SEVORANE AF*. If you have additional questions regarding SEVORANE AF*, please contact Abbott Laboratories, Limited.

Sincerely,

original signed by

Mauricio Ede, MD, PhD

Medical Director

Abbott Laboratories, Limited

*TM

Abbott Laboratories, Limited/Laboratoires Abbott, Limitée

P.O. Box/C.P. 6150, Station/succursale Centre-ville

Montréal (Québec) H3C 3K6

Any suspected adverse incident can also be reported to:

Health Products & Food Branch Inspectorate

HEALTH CANADA

Address Locator: 3002C

OTTAWA, Ontario, K1A 0K9

Tel: The Medical Devices Hotline 1-800-267-9675

Footnotes

Footnote 1

Holak EJ, Mei DA, Dunning MB III, et al. Carbon monoxide production from sevoflurane breakdown: modeling of exposures under clinical conditions. Anesth Analg. 2003;96:757-64.

Return to footnote 1 referrer

Footnote 2

Wissing H, Kuhn I, Warnken U, Dudziak R Carbon monoxide production from desflurane, enflurane, halothane, isoflurane, sevoflurane with dry sodalime. Anesthesiology 2001;95:1205-12.

Return to first footnote 2 referrer