ARCHITECT System - Creatine Kinase Assay (2018-07-20)
- Starting date:
- July 20, 2018
- Posting date:
- August 2, 2018
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-67350
Affected products
ARCHITECT System - Creatine Kinase Assay
Reason
The purpose of this is to inform customers of a product recall for the ARCHITECT Creatine Kinase (CK) reagent and to provide instructions on what actions must be taken. Abbott has identified a stability issue that may lead to the following issues.
- Error Code 1054 ""Unable to calculate results, Reaction Check Failure.
- Quality control (QC) results out of range low.
A potential delay may exist in the reporting of patient results due to quality control values falling outside of expected ranges and the generation of error code 1054. Internal testing on patient samples showed no atypical bias for the impacted reagent lots, therefore if quality controls are in range, patient results are not impacted.
Affected products
ARCHITECT System - Creatine Kinase Assay
Lot or serial number
- 24358UN18
- 87697UN18
- 99632UN18
Model or catalog number
7D63-21
Companies
- Manufacturer
-
Abbott Laboratories Diagnostcis Division
100 Abbott Park Road
Abbott Park
60064
UNITED STATES