Health product recall

ARCHITECT System - Creatine Kinase Assay (2018-07-20)

Starting date:
July 20, 2018
Posting date:
August 2, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-67350

Affected products

ARCHITECT System - Creatine Kinase Assay

Reason

The purpose of this is to inform customers of a product recall for the ARCHITECT Creatine Kinase (CK) reagent and to provide instructions on what actions must be taken. Abbott has identified a stability issue that may lead to the following issues.

  • Error Code 1054 ""Unable to calculate results, Reaction Check Failure.
  • Quality control (QC) results out of range low.

A potential delay may exist in the reporting of patient results due to quality control values falling outside of expected ranges and the generation of error code 1054. Internal testing on patient samples showed no atypical bias for the impacted reagent lots, therefore if quality controls are in range, patient results are not impacted.

Affected products

ARCHITECT System -  Creatine Kinase Assay

Lot or serial number
  • 24358UN18
  • 87697UN18
  • 99632UN18
Model or catalog number

7D63-21

Companies
Manufacturer

Abbott Laboratories Diagnostcis Division

100 Abbott Park Road

Abbott Park

60064

UNITED STATES