Health product recall

ARCHITECT BNP Controls (2019-05-29)

Starting date:
May 29, 2019
Posting date:
June 14, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-70199


Last updated:

Affected Products

ARCHITECT BNP Controls

Reason

Falsely elevated patient and/or control results may be obtained when using ARCHITECT BNP calibrators and controls. Preliminary investigation has determined that all current in-date BNP calibrator and control lots are impacted by an age dependent shift in performance. Use of impacted calibrator lots may result in falsely elevated patient results. Use of impacted control lots may result in invalid patient results due to controls out of range.

Affected products

ARCHITECT BNP Controls

Lot or serial number
  • 44K78918
  • 44K79318
  • 44K80818
  • 44K80918
  • 44K82118
  • 44K82618
  • 44K82718
  • 44K85119
Model or catalog number
  • 8K28-02
  • 8K28-11
Companies
Manufacturer

Abbott Laboratories Diagnostics Division

100 ABBOTT PARK ROAD

ABBOTT PARK

60064

Illinois

UNITED STATES