Health product recall

ARCHITECT BNP Calibrators and Controls (2019-05-29)

Starting date:
May 29, 2019
Posting date:
June 14, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-70203


Last updated:

Affected Products

A. ARCHITECT BNP Calibrators

B. ARCHITECT BNP Controls

Reason

Performance. Use of impacted calibrator lots may result in falsely elevated patient results. Use of impacted control lots may result in

invalid patient results due to controls out of range.

Affected products

A. ARCHITECT BNP Calibrators

Lot or serial number
  • 44K79418
  • 44K79518
  • 44K81018
  • 44K81118
  • 44K82018
  • 44K82818
  • 44K84119
  • 44K84219
  • 44K85219
Model or catalog number

8K28-03

Companies
Manufacturer

Abbott GMBH & Co. K.G.

MAX PLANCK RING 2

WIESBADEN

65205

GERMANY



B. ARCHITECT BNP Controls

Lot or serial number

44K79518

44K81118

44K82818

44K84219

Model or catalog number

8K28-12

Companies
Manufacturer

Abbott GMBH & Co. K.G.

MAX PLANCK RING 2

WIESBADEN

65205

GERMANY