This page has been archived on the Web
Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.
Arch Products
- Starting date:
- October 31, 2012
- Posting date:
- December 10, 2012
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public (GP), Healthcare Professionals, Hospitals
- Identification number:
- RA-16287
Recalled Products
- Rumi Arch (UMH700)
- Advincula Arch (UMH750)
Reason
Excessive pressure exerted on the uterine tip when attached to the arch may, under certain circumstances, cause the internal steel rod of the tip to protrude through the cleaning hole on the shaft of the handle.
Affected products
A. Rumi Arch (UMH700)
Lot or serial number
More than 1000 numbers, contact manufacturer
Model or catalog number
UMH700
Companies
- Manufacturer
- Coopersurgical Inc.
B. Advincula Arch (UMH750)
Lot or serial number
More than 1000 numbers, contact manufacturer
Model or catalog number
UMH750
Companies
- Manufacturer
- Coopersurgical Inc.