This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Aquios Cl Flow Cytometer

Starting date:
September 5, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-60270

Affected Products

Aquios Cl Flow Cytometer

Reason

The use of positions 9 and 10 of the reagent carousel can lead to misalignment of the carousel and damage to the prep probe. Damage to the prep probe will render the instrument inoperative with the potential to delay reporting of results.

Affected products

Aquios Cl Flow Cytometer

Lot or serial number

AY03004

AY25038

AZ12017

Model or catalog number

B30166

Companies
Manufacturer

Beckman Coulter Ireland, Inc.

Lismeehan, O'Callaghan's Mills

County Clare

IRELAND