Aquios Cl Flow Cytometer

Starting date:
September 5, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-60270

Affected Products

Aquios Cl Flow Cytometer

Reason

The use of positions 9 and 10 of the reagent carousel can lead to misalignment of the carousel and damage to the prep probe. Damage to the prep probe will render the instrument inoperative with the potential to delay reporting of results.

Affected products

Aquios Cl Flow Cytometer

Lot or serial number

AY03004
AY25038
AZ12017

Model or catalog number

B30166

Companies
Manufacturer
Beckman Coulter Ireland, Inc.
Lismeehan, O'Callaghan's Mills
County Clare
IRELAND