APOLLO AND JUNO DRF (2016-02-17)
- Starting date:
- February 17, 2016
- Posting date:
- April 4, 2016
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-57736
Affected products
- APOLLO-MAIN UNIT
- JUNO DRF
Reason
An operator got hurt while relocating a patient from table onto the patient's bed. Standing behind the table top between the left table top support arm and the detector holder, he got squeezed between the left table top support arm and the detector holder.
Affected products
-
APOLLO-MAIN UNIT
Lot or serial number
- S/N 08060167
- S/N 08070501
Model or catalog number
- 9784AXBYVT
Companies
- Manufacturer
-
Villa Sistemi Medicali S.P.A.
via Delle Azalee 3
Buccinasco
20090
ITALY
B. JUNO DRF
Lot or serial number
All
Model or catalog number
9784729831
Companies
- Manufacturer
-
Villa Sistemi Medicali S.P.A.
via Delle Azalee 3
Buccinasco
20090
ITALY