Health product recall

Apollo

Starting date:
December 27, 2016
Posting date:
January 25, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-61972

Reason

Villa Sistemi Medicali became aware that a patient fell from the table because the footrest of the device detached from the table top while the patient was positioned on the table and the table was moving from vertical position.

The analysis showed that the footrest of the device detached from the table top because the safety screws were discarded at installation. The purpose of these screws is to prevent the footrest to completely detach from the table in order to avoid the patient to fall.

Affected products

Apollo

Lot or serial number

  • 08060167
  • 08070501

Model or catalog number

  • 9784020231
  • 9784120231

Companies

Manufacturer

Villa Sistemi Medicali S.P.A.

via Delle Azalee 3

Buccinasco

20090

ITALY