Apollo
- Starting date:
- December 27, 2016
- Posting date:
- January 25, 2017
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-61972
Reason
Villa Sistemi Medicali became aware that a patient fell from the table because the footrest of the device detached from the table top while the patient was positioned on the table and the table was moving from vertical position.
The analysis showed that the footrest of the device detached from the table top because the safety screws were discarded at installation. The purpose of these screws is to prevent the footrest to completely detach from the table in order to avoid the patient to fall.
Affected products
Apollo
Lot or serial number
- 08060167
- 08070501
Model or catalog number
- 9784020231
- 9784120231
Companies
- Manufacturer
-
Villa Sistemi Medicali S.P.A.
via Delle Azalee 3
Buccinasco
20090
ITALY