Health product recall

ANTI-CD30 (BER-H2) MOUSE MONOCLONAL PRIMARY ANTIB (2015-08-05)

Starting date:
August 5, 2015
Posting date:
September 1, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-54860

Affected Products

  1. ANTI-CD30 (BER-H2) MOUSE MONOCLONAL PRIMARY ANTIB

Reason

A slight decrease of the staining intensity and percent positive cells was observed. No cases demonstrating complete negative (false negative results) were observed. The investigation results support that the expiration date for the affected lot numbers should be set to 18 months from the date of manufacture instead of the currently marked 24 months.

Affected products

A. ANTI-CD30 (BER-H2) MOUSE MONOCLONAL PRIMARY ANTIB

Lot or serial number
  • D09123
  • E00209
  • E03668
  • E05896
Model or catalog number
  • 790-4858