This page has been archived on the Web
Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.
Animas Insulin Infusion Pumps
- Starting date:
- December 17, 2012
- Posting date:
- January 21, 2013
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-16885
Recalled Products
- Animas IR 1200 Insulin Infusion Pump
- Animas 2020 Insulin Infusion Pump
- Animas 2020 Insulin Infusion Pump
Reason
The Animas IR 1200, after midnight on December 31, 2015, the internal calendar will revert back to a previous year. This pump model will continue to deliver insulin as programmed by the use but the history will record the insulin delivery in the incorrect year.
Affected products
A. Animas IR 1200 Insulin Infusion Pump
Lot or serial number
All Lots.
Model or catalog number
More than 10 numbers, contact the manufacturer.
Companies
- Manufacturer
-
Animas Corporation
B. Animas 2020 Insulin Infusion Pump
Lot or serial number
All Lots.
Model or catalog number
- 200-260-51
- 200-261-51
- 200-262-51
- 200-263-51
- 200-264-51
- 200-265-51
- 200-370-00
- 200-372-00
- 200-374-00
- 200-375-00
Companies
- Manufacturer
-
Animas Corporation
C. Animas 2020 Insulin Infusion Pump
Lot or serial number
All Lots.
Model or catalog number
- 100-380-51
- 100-381-51
- 100-382-51
- 100-384-51
- 100-385-51
- 200-370-51
- 200-371-51
- 200-372-51
- 200-374-51
- 200-375-51
Companies
- Manufacturer
-
Animas Corporation