and Infinity M540 IACS Monitoring Solution With C700 and Infinity P2500 (2019-04-22)
- Starting date:
- April 22, 2019
- Posting date:
- January 10, 2020
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-72061
Last updated: 2020-01-10
Affected Products
- IACS Monitoring Solution With C700
- Infinity P2500
- Infinity M540
Reason
Immediately after the infinity M540 patient monitor has been docked to the M500 docking station and a user discharges a patient from the M540 patient monitor, there is a possibility that the communication between the IACS cockpit and the M540 patient monitor is not established. If the communication is not established, the IACS cockpit will not be able to send any commands to the M540 patient monitor with the exception of the "audio pause". Changes made on the IACS cockpit will not be transferred to the M540 patient monitor, though the changes will be reflected on the IACS cockpit.
Affected products
-
IACS Monitoring Solution With C700
Lot or serial number
- 5613496066
- 5613500775
- 5613501373
- 5613503175
- 5613505173
- 5613505369
- 5613505966
- 5613516865
Model or catalog number
MS25510
Companies
- Manufacturer
-
Draeger Medical Systems, Inc.
3135 Quarry Road
Telford
18969
Pennsylvania
UNITED STATES
B. Infinity P2500
Lot or serial number
USLM-0689
USLM-0790
Model or catalog number
MS25520
MS22277-XX
Companies
- Manufacturer
-
Draeger Medical Systems, Inc.
3135 Quarry Road
Telford
18969
Pennsylvania
UNITED STATES
C. Infinity M540
Lot or serial number
5613503371
5613507866
5613508767
5613513369
Model or catalog number
MS20401
MS26372
Companies
- Manufacturer
-
Draeger Medical Systems, Inc.
3135 Quarry Road
Telford
18969
Pennsylvania
UNITED STATES