Amia Automated PD System - Cycler (2018-07-15)
- Starting date:
- July 15, 2018
- Posting date:
- January 31, 2019
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-68952
Affected Products
Amia Automated PD System - Cycler
Reason
Baxter Corporation has identified a potential failure with the Amia cycler where the power cord socket can become loose or detached when the power cord is unplugged from the back of the cycler. To address this issue, Baxter will be inspecting and repairing any loose power cord sockets during the next Amia cycler service. The Amia cyclers may continue to be used, and do not need to be returned at this time.
Affected products
Amia Automated PD System - Cycler
Lot or serial number
More than 100 numbers, contact manufacturer
Model or catalog number
5C9320
Companies
- Manufacturer
-
Baxter Healthcare Corporation
1 Baxter Parkway
Deerfield, Illinois
60015
UNITED STATES