Health product recall

Amia Automated PD System - Cycler (2018-07-15)

Starting date:
July 15, 2018
Posting date:
January 31, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-68952

Affected Products

Amia Automated PD System - Cycler

Reason

Baxter Corporation has identified a potential failure with the Amia cycler where the power cord socket can become loose or detached when the power cord is unplugged from the back of the cycler. To address this issue, Baxter will be inspecting and repairing any loose power cord sockets during the next Amia cycler service. The Amia cyclers may continue to be used, and do not need to be returned at this time.

Affected products

Amia Automated PD System - Cycler

Lot or serial number

More than 100 numbers, contact manufacturer

Model or catalog number

5C9320

Companies
Manufacturer

Baxter Healthcare Corporation

1 Baxter Parkway

Deerfield, Illinois

60015

UNITED STATES