Ambu Aview 2 Advance
Brand(s)
Last updated
Summary
Product
Ambu Aview 2 Advance
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products | Lot or serial number | Model or catalogue number |
---|---|---|
Ambu Aview 2 Advance | 054 to 056 & 058 to 059 | 405011000 |
Issue
Ambu have received input from market where Ambu® Aview 2 advance was dropped and upon impact hitting the floor the lithium-ion battery in the device short-circuited and combusted whereby smoke and flames was seen. No harm to patient, user or third party has been reported in any of the cases.
Recall start date: August 13, 2023
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies
Ambu A/S |
Baltorpbakken 13, Ballerup, Denmark, 2750 |
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-75068
Get notified