Health product recall

Altrix Precision Temperature Management System (2018-09-04)

Starting date:
September 4, 2018
Posting date:
September 28, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-67830

Affected products

Altrix Precision Temperature Management System

Reason

  1. Users may experience alarm fatigue due to frequent alarming relating to patient temperature deviation beyond 0.5 degrees Celsius while in automatic mode.
  2. Users may experience a remove from use code 9 (RFU 9) fault condition relating to inadvertent flow alarms.
  3. Users may experience code RFU 27fault condition relating to mechanical interference between the device fan and filter.

Affected products

Altrix Precision Temperature Management System

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

8001

Companies
Manufacturer
Stryker Medical
3800 E Centre Avenue
Portage
49002
Michigan
UNITED STATES