AlterG M320 (2018-02-09)
- Starting date:
- February 9, 2018
- Posting date:
- March 2, 2018
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-66072
Affected products
AlterG M320
Reason
A condition was reported by a distributor in which an M320 unit unexpectedly surged in speed to the maximum allowed by the treadmill. The treadmill e-stop was not effective. Failure traced to a specific diode on the lower motor controller.
Affected products
AlterG M320
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
ALTERG M320
Companies
- Manufacturer
-
AlterG Inc.
48368 Milmont Dr
Fremont
94538
California
UNITED STATES