Health product recall

AlterG M320 (2018-02-09)

Starting date:
February 9, 2018
Posting date:
March 2, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-66072

Affected products

AlterG M320

Reason

A condition was reported by a distributor in which an M320 unit unexpectedly surged in speed to the maximum allowed by the treadmill. The treadmill e-stop was not effective. Failure traced to a specific diode on the lower motor controller.

Affected products

AlterG M320

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

ALTERG M320

Companies
Manufacturer
AlterG Inc.
48368 Milmont Dr
Fremont
94538
California
UNITED STATES